Журнал Российского общества по неразрушающему контролю и технической диагностике
The journal of the Russian society for non-destructive testing and technical diagnostic
 
| Русский Русский | English English |
 
Главная Archive
22 | 12 | 2024
2019, 09 September

DOI: 10.14489/td.2019.09.pp.058-063

 

Lelyukhin A. S., Muslimov D. A., Piskaryova T. I., Shekhtman A. G.
METHODS FOR DETERMINING PARAMETERS OF RADIOPAQUE INCLUSIONS BY ABSORPTION CURVES
(pp. 58-63)

Abstract. The method for determining the effective atomic number and surface density of filters based on x-ray absorption data in a linear multichannel detector material is considered in this paper. Data on the absorption of radiation is proposed to consider as the coordinates of vectors in the spectrozonal space. It is shown that the relative change in length and the angle of divergence of the vectors can be used as criteria for the calibration of the spectrozonal recording system. Calibration characteristics are obtained for the mathematical model of the spectrozonal recording system. Calibration was performed for filters from materials with a priori given effective atomic numbers from 6 to 40 with a surface density of from 0.05 to 1.5 g/cm2. It has been established that in the calibration range, the deviations of the measured values of the parameters of the filters do not exceed 10 % by the effective atomic number and 30 % by the surface density. The proposed measurement method can be used to quantify the mineralization of the charge materials, in the separation of minerals, or to search for prohibited inclusions in covering environments. The efficiency of the technique can be achieved by optimizing the dimensionality of the spectrozonal space and choosing the conditions for the excitation of the primary radiation beam.

Keywords: X-rays, absorption curves, effective atomic number, surface density.

A. S. Lelyukhin, D. A. Muslimov, T. I. Piskaryova (Orenburg State University, Orenburg, Russia) E-mail: Данный адрес e-mail защищен от спам-ботов, Вам необходимо включить Javascript для его просмотра. , Данный адрес e-mail защищен от спам-ботов, Вам необходимо включить Javascript для его просмотра. , Данный адрес e-mail защищен от спам-ботов, Вам необходимо включить Javascript для его просмотра.
A. G. Shekhtman (Orenburg State Medical University, Orenburg, Russia) E-mail: Данный адрес e-mail защищен от спам-ботов, Вам необходимо включить Javascript для его просмотра.

 

1. Rebuffel V., Dinten J. M. (2007). Dual-energy X-ray imaging: benefits and limits. Insight-non-destructive testing and condition monitoring, 49(10), pp. 589 – 594.
2. Rühl R., Wozniak M. M., Werk M. et al. (2008). CsI-detector-based dual-exposure dual energy in chest radiography for lung nodule detection: results of an international multicenter trial. European radiology, 18(9), pp. 1831.
3. Taibi A., Fabbri S., Baldelli P. et al. (2003). Dual-energy imaging in full-field digital mammography: a phantom study. Physics in Medicine & Biology, 48(13), p. 1945.
4. Ryzhikov V. D., Naydenov S. V., Grinyov B. V. et al. (2003). Multi-energy radiography on the basis of “scintillator–photodiode” detectors. Nuclear Instruments and Methods in Physics Research. Section A. Accelerators, Spectrometers, Detectors and Associated Equipment, 505(1 – 2), pp. 549 – 551.
5. Nittoh K., Sakurai T., Oyazu E. et al. (2003). Extension of dynamic range in X-ray radiography using multi-color scintillation detector. Nuclear Instruments and Methods in Physics Research Section A. Accelerators, Spectrometers, Detectors and Associated Equipment, 501(2 – 3), pp. 615 – 622.
6. Ryzhikov V. D., Opolonin O. D., Naydenov S. V. et al. (2005). Study of a dual-energy line of detectors for x-ray osteodensitometry. Meditsinskaya tekhnika, (2), pp. 18 – 21. [in Russian language]
7. Blake G. M., Fogelman I. (1997). Technical principles of dual energy x-ray absorptiometry. Seminars in nuclear medicine. WB Saunders, 27(3), pp. 210 – 228.
8. Chun K. J. (2011). Bone densitometry. Seminars in Nuclear Medicine. WB Saunders, 41(3), pp. 220 – 228.
9. Lelyuhin A. S. (2017). Determination of the quality parameters of X-ray beams from absorption curves. Meditsinskaya tekhnika, (3), pp. 35 – 39. [in Russian language]
10. Lelyuhin A. S., Piskareva T. I., Kornev E. A. (2018). Non-classical X-ray spectrometer based on a linear multichannel detector. Prikladnaya fizika, (2), pp. 90 – 96. [in Russian language]
11. Lelyuhin A. S., Muslimov D. A., Taisov M. V., Adzhieva M. D. (2014). A set of software modules for restoring the bremsstrahlung spectrum by minimizing directional discrepancy. Certificate on state registration of a computer program No. 2014611219. Federal Service for Intellectual Property. Moscow. [in Russian language]
12. Lelyukhin A. S., Kornev E. A., Kan’shin V. V. (2004). Spectrozonal X-ray detector for determining the effective atomic number of absorbing tissue. Biomedical Engineering, 38(5), pp. 227 – 232.

 

 

This article  is available in electronic format (PDF).

The cost of a single article is 350 rubles. (including VAT 18%). After you place an order within a few days, you will receive following documents to your specified e-mail: account on payment and receipt to pay in the bank.

After depositing your payment on our bank account we send you file of the article by e-mail.

To order articles please copy the article doi:

10.14489/td.2019.09.pp.058-063

and fill out the  form  

 

 

 
Search
Rambler's Top100 Яндекс цитирования