Журнал Российского общества по неразрушающему контролю и технической диагностике
The journal of the Russian society for non-destructive testing and technical diagnostic
 
| Русский Русский | English English |
 
Главная Archive
22 | 12 | 2024
2015, 05 May

DOI: 10.14489/td.2015.05.pp.026-031

 

Chuprin V.A.
INVESTIGATION OF THE POSSIBILITY OF AUTOMATIC MOTOR OILS PARAMETER MEASUREMENTS WITH ULTRASONIC VISCODENSITOMETER
(pp. 26-31)

Abstract. The article focuses on the application of a new class of devices – ultrasonic viscodensitometer to measure the viscosity and density of automotive engine oils. The objects of study are motor oils SAE 0W20, 10W30, 5W40, 10W40, 15W40, 20W50. The results of laboratory measurements are compared with the results of measuring the density and viscosity of engine oils by ultrasonic methods.Researches show that the ultrasonic viscodensitometer allows to measure accurately viscosity and density of motor oils. However, measurements of the oil features by ultrasonic methods should be taken into consideration the fact that motor oils are non-Newtonian fluids as well as the corrective additives in the composition of oils can introduce significant errors. Therefore, calibration of a device requires a cali-bration curve with use up to three reference liquids.

Keywords: motor oil, viscosity, density, horizontally-polarized normal wave, symmetric Lamb wave, ultrasonic viscodensitometer.

 

V. A. Chuprin
SIC “LUCH” Ltd. Balashiha, Moscow region, Russia. E-mail: Данный адрес e-mail защищен от спам-ботов, Вам необходимо включить Javascript для его просмотра.

 

 

1. Kovalenko V. P. (1987). Pollution and purification of petroleum oils. Moscow: Khimia.
2. Zhu J., He D., Bechhoefer E. (2013). Survey of lu-brication oil condition monitoring, diagnostics, and prognos-tics techniques and systems. Journal of Chemical Science and Technology, 2(3), pp. 100-115.
3. Agoston A, Ötsch C., Jakoby B. (2005). Viscosity sensors for engine oil condition monitoring–application and interpretation of results. Sensors and Actuators, (A 121), pp. 327-332. doi: 10.1016/j.sna.2005.02.024.
4. Chuprin V., Gitis M. (2012). Measurements of the material properties of liquids using normal acoustic plate waves. Proceedings 18th World NDT Conference. Durban, 2012.
5. Chuprin V. A. (2013). The fluid influence on the normal waves parameters in a plate waveguide, immersed in liquid. Kontrol'. Diagnostika, (10), pp. 71-75.
6. Deshpande A. P., Krishnan J. M., Kumar S. (Eds.). (2010). Rheology of complex fluids. Berlin: Springer.

 

 

This article  is available in electronic format (PDF).

The cost of a single article is 350 rubles. (including VAT 18%). After you place an order within a few days, you will receive following documents to your specified e-mail: account on payment and receipt to pay in the bank.

After depositing your payment on our bank account we send you file of the article by e-mail.

To order articles please fill out the form below:

Purchase digital version of a single article


Type the characters you see in the picture below



 

 

 
Search
Rambler's Top100 Яндекс цитирования