Журнал Российского общества по неразрушающему контролю и технической диагностике
The journal of the Russian society for non-destructive testing and technical diagnostic
 
| Русский Русский | English English |
 
Главная
23 | 12 | 2024
2017, 06 June

DOI: 10.14489/td.2017.06.pp.042-049

 

Golobokov M. V., Danilevich S. B.
DETERMINATION OF THE VALUE OF SIGHT AND SIZE SOURCE EFFECT FOR CALIBRATION OF INFRARED THERMOMETERS AND THERMAL IMAGERS
(pp. 42-49)

Abstract. The paper gives the analysis of need and sufficiency of requirements of techniques of checking the infrared thermometers and thermal imagers regulating measure definition of sighting and intrinsic error. Recommendations for improving the effectiveness of calibration of measuring instruments are proposed. The requirements of methods of infrared thermometers verification in terms of determining the indicator of sight are analyzed. The necessity to make calculations by determination of the sighting of energy values is proved. The influence of temperature of blackbody model on the accuracy of the calibration results are analyzed. The values of the absolute blackbody temperature are given for these infrared thermometers models under consideration, which are likely to calculate in terms of temperature. The ambiguity of requirements of verification methods in terms of working distance and aperture of used blackbody models are analyzed by the example of infrared thermometers and thermal imagers. The influence of the source size effect on the measurement result and as a result on the reliability of the verification results is shown. The problems of comparability of measurement results obtained with use of models blackbody in the form of emitting surface and a heat pipe are considered. The necessity of taking into account the distance and aperture of used blackbody model at normalized conditions are justified.

Keywords: infrared thermometer, thermal imager, sighting index, effect of the size of a source.

 

M. V. Golobokov (FSAEI DPO «Novosibirsk Branch of the Academy of Standardization, Metrology, and Certification (Training)», Novosibirsk, Russia) E-mail: Данный адрес e-mail защищен от спам-ботов, Вам необходимо включить Javascript для его просмотра.
S. B. Danilevich (State Regional Center of Standartization, Metrology and Testing in the Novosibirsk Region, Novosibirsk, Russia) E-mail: Данный адрес e-mail защищен от спам-ботов, Вам необходимо включить Javascript для его просмотра.

 

 

1. Frunze A.V. (2012). The methodical error of power pyrometers caused by influence on result of measurement of distance between a pyrometer and an object. Izmeritel'naya tekhnika, (10), pp. 37-41. [in Russian language]
2. Infrared thermometer of AKIP series. (2008). Verification procedure No. MP 40283–08. Moscow: Izdatel'stvo VNIIMS. [in Russian language]
3. Infrared thermometer of AKIP series. (2013). Verification procedure No. MP 53856-13. Moscow: Izdatel'stvo VNIIMS. [in Russian language]
4. Verification procedure of infrared thermometer Fluke 572, 574, 576 series. (2015). Moscow: Izdatel'stvo Rostest-Moskva. [in Russian language]
5. Gaussorgues G. (1984). Infrared thermography. Fundamentals, techniques, application. Moscow: Mir. [in Russian language]
6. Minkina W., Dudzik S. (2009). Infrared thermography. Errors and uncertainties. N.Y.: John Wiley & Sons.
7. Radiation thermometers «Raynger». (2007). Verification procedure No. MP 2412-0016-2007. St. Petersburg: Izdatel'stvo VNIIM im. D. I. Mendeleeva. [in Russian language]
8. Infrared thermometer of «SIGHT» series of modifications MS, MSPlus, MSPro, LS, LSDCI, P20LT, P20 1M, P20 2M. Verification procedure. (2010). St. Peterburg: Izdatel'stvo VNIIM im. D. I. Mendeleeva. [in Russian language]
9. Converters primary pyrometric of complete and partial radiation. Verification procedure. (1986). Study guide No. MI 1200-86. Moscow: Izdatel'stvo Standartinform. [in Russian language]
10. Liebmann. F. (2010). Use of new standards for hand-held infrared thermometer calibration. NCSL International Workshop and Symposium.
11. Moiseeva N.P. Verification of radiation thermometers. Available at: http://temperatures.ru/pages/ poverka_radiacionnyh_termometrov (Accessed: 23 June 2016) [in Russian language]
12. State system for ensuring the uniformity of measurements. Thermographic instruments. Verification procedure. (2006). Ru Standard No. GOST R 8.619-2006. Moscow: Izdatel'stvo Standartinform. [in Russian language]
13. Infrared camera FlexCam, InSight Т. Verification procedure. (2005). Moscow: Izdatel'stvo VNIIOFI. [in Russian language]
14. Infrared camera. Verification procedure. (2006). Moscow: Izdatel'stvo VNIIOFI. [in Russian language]
15. Infrared camera FLIR T 335, FLIR T 365, FLIR T 425, FLIR В 335, FLIR В 365, FLIR В 425, FLIR i7 made by «FLIR Systems AB», Sweden. (2010). Moscow: Izdatel'stvo VNIIOFI. [in Russian language]
16. Thermal imagers infrared «AGEMA 570» made by «FSI, FlirSystems AV» (Sweden). Verification procedure. (1999). Moscow: Izdatel'stvo Rostest-Moskva. [in Russian language]
17. Thermal imagers infrared «THERMACAM PM 675, PM 695, SC 300, SC 500, SC 3000». Verification procedure. (2001). Moscow: Izdatel'stvo Rostest-Moskva. [in Russian language]
18. ThermaCAM P60, P40, S40, S60 and E2, E1, EM. Verification procedure. (2003). Moscow: Izdatel'stvo Rostest-Moskva. [in Russian language]
19. Golobokov M.V. Danilevich S.B. (2016). A simulation model of measuring instruments calibration procedure. Kompetentnost', (4), pp.40-47. [in Russian language]

 

 

This article  is available in electronic format (PDF).

The cost of a single article is 350 rubles. (including VAT 18%). After you place an order within a few days, you will receive following documents to your specified e-mail: account on payment and receipt to pay in the bank.

After depositing your payment on our bank account we send you file of the article by e-mail.

To order articles please fill out the form below:

Purchase digital version of a single article


Type the characters you see in the picture below



 

 

 
Rambler's Top100 Яндекс цитирования